8. Insert the re-wetted transducer capsule into the transducer housing inside the
meter body.
9. Slide the spring and shoulder washer along the cable to the transducer capsule.
Install the retaining ring into the groove within the transducer housing while
compressing the spring by applying force to shoulder washer.
Figure 3-19: Reassemble transducer components
A. Feed thru cable seal
B. Retaining ring
C. Shoulder washer
D. Spring
E. Transducer capsule
F. Transducer housing
10. Slide the feed-thru cable seal along the cable towards the transducer housing.
Insert the feed thru cable seal into the housing groove with the retaining ring pliers.
11. Repeat
for re-wetting the other pair of transducers on the
meter.
12. Insert the feed-thru cable seal into the housing groove with the Retaining Ring
Pliers.
13. Apply electrical power to the meter.
14. Connect to the meter with MeterLink and run the Zero Flow Calibration Wizard.
Then, collect and save a Maintenance log to show the “as left” condition of the
meter. Or use AMS
™
Device Manager or a HART
®
Field Communicator and select
Overview
→
Zero Calibration
→
Start menu
or
Service Tools
→
Maintenance
→
Zero Calibration
→
Start
.
15. Reinstall the shroud. To do so, go to:
Maintenance and Troubleshooting Manual
Meter maintenance
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
55
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...