Figure 3-59: Backplane board replacement
A. Non-terminal end of Transmitter Electronics Enclosure
B. Power Supply board
C. I.S. Barrier board (inside the Guide Plate)
D. Backplane board
7. Use a 1/8 inch (3 mm) flat head screw driver and disconnect the Acquisition Cable
terminal block from the Backplane. Unplug the Acquisition Cable from the
Backplane.
8. Remove the Power Supply (if it was not removed with the Backplane board) and I.S.
Barrier boards from the enclosure. The I.S. Barrier Board has a notched tab that
secures the board to the Guide Plate.
9. Attach the Acquisition Cable terminal block to the new Backplane Board and plug
the Power Supply Board and I.S. Barrier board into the Backplane board.
10. Insert the Backplane (with the Power Supply and I.S. Barrier Boards attached to the
Backplane) into the enclosure.
11. Fully seat the CPU Module and Optional I/O Module onto the Backplane board.
12. Install the four Phillips head screws to secure the Backplane to the enclosure
standoffs.
13. Reinstall the terminal blocks on the CPU Module, Optional I/O Module, and the
Power Supply board using a 3 mm flat head screw driver.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
96
Emerson.com/Rosemount
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...