NOTICE
Make sure the terminal blocks are aligned with the Guide Plate openings.
14. Recheck the connections, wiring and switch settings before replacing the end caps.
15. If replacing other electronics, continue with the following sections before replacing
the end caps and sealing the enclosure.
16. If you are not replacing other electronics, replace the end caps, security latches,
reseal the meter and apply power. If required, install the security seals through the
hex bolts in the Junction box cover (see
Seal the Junction Box for remote mount
).
17. Apply power to the meter.
18. Connect to the meter with MeterLink. Open the Meter Monitor (Detailed) view and
verify the meter is acquiring data, the transducers have good signals and flow
profiles for the chords are displayed and run the Zero Flow Calibration Wizard;
select the
Calibrate
→
Zero Calibrate
menu and generate the report. Then, collect
and save a Maintenance log to show the “as left” condition of the meter. Or use
AMS
™
Device Manager or a HART
®
Field Communicator and select
Overview
→
Zero Calibration
→
Start
menu or
Service Tools
→
Maintenance
→
Zero
Calibration
→
Start
.
19. Close your application connection (MeterLink, AMS
™
Device Manager or HART
®
Field Communicator).
20. This completes the Backplane board replacement procedure.
If you encounter problems with this procedure, see the Emerson Customer Support
contact information on the back cover of this manual.
3.9.4
Replace the I.S. Barrier board
Procedure
1. If replacing the I.S. Barrier board, remove power to the meter.
2. Refer to
for the tools required to complete this procedure.
3. Disconnect the Transmitter Electronics Enclosure security seals, loosen the end cap
security latches with a 3mm Allen wrench and remove both end caps (see
).
4. Use a 1/8 inch (3 mm) flat head screw driver and remove the terminal blocks from
the Power Supply board, the CPU Module and the Optional I/O Module. See
for board locations and associated terminal blocks.
5. Use a Phillips head screw driver and remove the four Backplane board screws from
the enclosure standoffs.
6. Pull the Backplane board out of the enclosure. This disconnects the I.S. Barrier
Board. Lay the Backplane board down with the Acquisition Cable still attached (the
Power Supply board may remain attached to the Backplane when you remove it
from the enclosure).
Maintenance and Troubleshooting Manual
Meter maintenance
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
97
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...