Table 2-2: Maintenance
(continued)
MeterLink utility
Diagnostics
Action(s)
Meter Monitor (Detailed) view
Meter Data List
Electronics temperature out of
range
Temperature of the electronics is out of
nominal operating range below -40°C or above
100°C (-40°F or above 212°F).
•
Heat or cool the meter electronics housing.
If operating temperature exceeds 65°C,
remote mount the Transmitter Electronics
Enclosure.
Meter Monitor (Detailed) view
Meter Data List
Electronics voltage out of
range
•
CPU board system voltages are valid if 1.0V,
1.2V, 2.5V, 3.3V or the Acquisition Module
valid voltages are 1.2V, 2.5V or 3.3V.
•
Replace the CPU Module if one or more of
the System Voltages is out of range.
•
Replace the Acquisition Module if one or
more of the voltages is out of range.
MeterLink Tools Menu
Frequency output
•
Run the Frequency Outputs test.
•
If the output reads zero, you may require a
pull up resistor 1.2kOHM, 0.5W.
•
Check frequency output from minimum to
maximum values.
Maintenance and Troubleshooting Manual
Troubleshooting
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
27
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...