21. Reseal the meter electronics and shroud, if required (see
).
This completes the transducer housing and seal replacement procedure.
3.7
Transducers replacement
This section discusses replacing the transducers for direct mount and remote mount
electronics options specific to your meter size (2” - 3”, 4” - 10”, and 12” and larger
meters).
Tip
Liquid transducer capsules can be replaced a single transducer capsule at a time. They do
not need to be replaced in pairs. Average delay and delta times are based on the
transducer housing so they will not change in the meter configuration when transducer
capsules are replaced.
WARNING
CUTTING HAZARD
Sharp edges may be present on the meter.
Wear appropriate personal protective equipment when working on the meter.
Failure to comply may cause serious injury.
WARNING
FLUID CONTENTS MAY BE UNDER PRESSURE
When the meter is under pressure, DO NOT attempt to remove or adjust the transducer
housing.
Attempting to do so may release pressurized fluid, resulting in serious injury for personnel
or equipment damage.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
62
Emerson.com/Rosemount
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...