Figure 3-23: Reassemble transducer components
A. Feed thru cable seal
B. Retaining ring
C. Shoulder washer
D. Spring
E. Transducer capsule
F. Transducer housing
15. Slide the feed-thru cable seal along the cable towards the transducer housing.
Insert the feed thru cable seal into the housing groove with the Retaining Ring
pliers.
16. Repeat
for the second transducer housing.
17. Apply electrical power to the meter.
18. Connect to the meter with MeterLink and run the Zero Flow Calibration Wizard;
select the
Calibrate
→
Zero Calibrate
menu and generate the report. Then, collect
and save a Maintenance log to show the “as left” condition of the meter. Or use
AMSTM Device Manager or a HART
®
Field Communicator and select
Overview
→
Zero Calibration
→
Start
menu or
Service Tools
→
Maintenance
→
Zero
Calibration
→
Start
.
19. Close your application connection (MeterLink, AMS
™
Device Manager or HART
®
Field Communicator).
20. Reinstall the shroud:
•
•
•
Maintenance and Troubleshooting Manual
Meter maintenance
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
61
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...