Table 3-7: Transducer removal tool list
(continued)
Tools required
MeterLink, or AMS
™
Device Manager or a HART
®
Field Communicator
Procedure
1. Disable electrical power to the meter.
2. Disassemble the shrouds. Refer to the appropriate sections below for your meter:
•
•
•
•
3. If the installation has rigid conduit, use a medium size crescent wrench and loosen
the hex nuts on the Transmitter Electronics Enclosure. This should allow enough
slack to remove the Transmitter Electronics Enclosure from the Base Enclosure. If
the installation uses flexible conduit, you may not need to disconnect it from the
Transmitter Electronics Enclosure.
Figure 3-24: Conduit removal
A. Transmitter electronics enclosure
B. Conduit gland fitting
4. Use a 7/16 inch (11mm) wrench and remove the four hex head bolts and split lock
washers securing the Transmitter Electronics Enclosure to the Base Enclosure.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
64
Emerson.com/Rosemount