Figure 3-2: Remote mount transmitter electronics options
A. Bolted band shroud design 2" & 3" sizes
B. Split shroud design 4" to 10" - 150 & 300 ANSI sizes
C. Clamped band shroud design 12" and larger sizes
Transducer field removal and installation is required if a transducer is fouled or if a chord
failure occurs.
The LT-10 and LT-11 transducers have a spring loaded assembly with the piezoelectric
element at one end and the transducer cable at the other end. Meters are supplied with
transducers which are extractable while the line is pressurized. The transducer assembly
shown below is a one-piece capsule that can be easily installed or removed from the meter
without depressurizing the unit using the tool list provided in
.
•
LT-10 transducers are designed for 2 through 10 inch meter sizes.
•
LT-11 transducers are designed for 12 inch and larger meters sizes.
Maintenance and Troubleshooting Manual
Meter maintenance
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
41