Figure 3-17: Transducer removal
A. Feed-thru cable seal
B. Shoulder washer
C. Spring
D. Transducer capsule
Figure 3-18: LT-10 Transducer and cable assembly removed from meter
6. Place one drop of Dow Corning 200 Silicon Oil
1-360-01-650) on the face of the transducer. Make sure the oil is evenly distributed
across the transducer face.
7. Wipe excess oil from the transducer sides.
(1)
Dow Corning 200 is a trademark of Dow Corning Corporation, U.S.A.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
54
Emerson.com/Rosemount