Dresser Model 10C25 Series K Installation, Operation, and Maintenance Manual
© 2018 Natural Gas Solutions North America, LLC
Page 51 of 74
3.
Verify you are in the
Other
category in the
Mfr Type
area on the upper part of the screen. Click
ADD
METER
.
4.
Fill in every box with data. The MMX prover leaves information in each box, so change the areas that
require updates. Click the
ENTER
button after each input to save the data in each box.
Table 13 shows an example of the type of data listed in each field.
Table 13: Example SNAP Prover Data Fields with MMX Platforms
Field
Example Data
Meter Name
10C25
Proof Open
1000
Proof Check
100
Proof Other
500
Rate Open
800
Rate Check
80
Rate Other
400
Revs Open
4
Revs Check
2
Revs Other
3
Exercise Revs
2
Jog Rate
100
Revs/Unit Volume
1
Max Diff Set Point
0.5
Tolerances Pos
1
Tolerances NEG
-1
Slope
1 1
Repair
2
Meter Size
10C
5.
After updating every field with the correct data, click
SAVE SPECS
.
6.
Verify the caption
SPECS
SAVED
displays in the blue caption box. If this caption does not display, the
meter file was not saved. Retry step 5 until the meter test file saves.
7.
When you exit the test setup screen, verify the meter test file is listed in the
OTHER
manufacturer
category.
8.
You can now install the meter on the test bench (refer to Section 13.2.3).
9.
Exit to the test screen, and continue the normal process for running a test (refer to Section 13.3).
Note:
For SNAP Provers, the 10C25 meter must be tested by using a Large Meter Kit for
flows greater than 500 cfh.