Dresser Model 10C25 Series K Installation, Operation, and Maintenance Manual
© 2018 Natural Gas Solutions North America, LLC
Page 48 of 74
5.
On the
Configuration
screen, click
Upload to unit
to upload the
setting changes to the meter and
clear the highlighting.
Figure 61: Click Upload to unit.
6.
Select the appropriate prove mode on the LCD screen on the meter (refer to Section 13.1.1 or Section
13.1.2).
13.1.1
Compensated Prove Mode Setting
Perform the following steps on the meter’s LCD screen:
1.
Swipe the magnet across the black dot on the
meter until the screen displays
PROV C.V
, and
then stop swiping.
Figure 62: PROV C.V LCD screen
2.
After five (5) seconds, the display changes to
PROVE I.C.V
.
Figure 63: PROVE I.C.V LCD screen
3.
Hold the magnet on the black dot for
approximately five (5) seconds until the
display changes to
PRVE_CO.R
.
The meter is now ready to be proved by using
the compensated volume output.
Figure 64: PRVE_CO.R LCD screen
4.
To exit Prove Mode, hold the magnet on the black dot for five (5) seconds.
Note:
The unit will automatically exit prove testing mode in thirty (30) minutes.