V1.0
20170825
61
5. Use
5.8.4 Paused and Resume staining program
Paused
Resume
Touch to select any program that is being
executed in the current progress list, and
the button “Paused” is activated.
Click the button “ Paused ” , and the
system displays the discontinuation
dialog box.
If the button “Normal” is pressed, the
program enters the loading station after
the current staining procedures are
completed.
If the button “Immediately” is pressed,
the staining step of the current program is
immediately discontinued and the system
will directly enters the unloading station.
The staining program in discontinued
status in the current progress list is
displayed in red background
This function is applicable to the program that requires rapid
execution, such as frozen dyeing.
The use of discontinuation is not recommended for some
stations, such as 1% hydrochloric-alcohol solution. Forced use
may result in impairment of specimens.
On the current progress interface, touch
to select the discontinued staining
program displayed with red background,
and the button
“
Continue
”
is activated.
Click the button
“
Resume
”
, and the system
displays a prompt dialog box. After
“
Confirmation
”
, the staining steps of this
program that are not completed continue
to be executed.
Fig.59