4. Installation
4.1 Site Requirements
•
The experimental workbench is solid and completely horizontal, with smooth
and flat surface. Its minimum width is 1m and minimum height is 0.8m.
•
The farthest distance from the corresponding water inlet / outlet on the
apparatus back to the faucet is 2.5m and to the waste pipe is 2m.
•
If an exhaust pipe operation is to be performed on the device, there is a need
to install a ventilation hood at the position less than 3.5 cm away from the
apparatus.
•
There should be enough space (about 0.9m) above the experimental
workbench so that the modules of the apparatus can be easily opened /
closed.
•
There is almost no vibration on the floor.
•
Maintain environmental temperature stable b15℃ and +40℃.
•
The maximum relative humidity is 80% (without condensation).
•
Do not place any vibration source that may induce vibration near the
apparatus.
4.2 Installation of the Apparatus
•
`Elevate the apparatus and hold the
transport position of the apparatus.
•
Place the apparatus at the proper
position of a workbench.
•
Check whether all accessories are
received by referring to the
configuration checklist.
The total weight of apparatus is
about 90 kg. Therefore, 4
people are needed for the
elevation or transportation of
the apparatus.
The chemicals used in slide stainer are inflammable and toxic. The apparatus
must be installed in a well-ventilated area, and shall be kept away from any fire
source.
The apparatus shall not be used in the room with an explosion hazard.