5. Use
Drawer open timeout: Set range from 1-120s, click + or - or enter a number
through the soft keyboard, and the user can set delayed alarm time for the overtime
of drawer opening. For example, if it is set to 5S, the system gives the voice prompt
“Please Close Drawer” after the drawer is opened for over 5s.
Add water to Rinse station: Set range from 0 ~ 120s, click + or - or enter a
number through the soft keyboard, and the user can set the time of continuous flow
of water washing station. For example, if it is set to 6S, water inflow is continued for
another 6s after the current washing process is completed.
Increase oven heating time: Set range from 0 ~ 10 minutes, Click + or - or
enter a number through the soft keyboard, and the user can set the time of
continuous heating of oven. If it is set to 1 minute, the heating is continued for 1
minute after the current oven working procedure is completed.
Start fans density data: Set range from 0 ~ 100 ppm, Touch and pull the
prompt bar slide block to a proper value and then release. Click the “Setting”
button, and the setting of post-filtration concentration monitoring is completed. For
example, it is set to 40 ppm, the system automatically starts the ventilation system
to perform filtering when detecting that gas concentration is greater than 40 ppm,
so that gas concentration value is reduced.
Oven station temp: Set the range from 0 ~ 70 ℃, touch and pull the prompt bar
slide block to a proper value and then release. Click the “Setting” button, and the
setting of oven temperature monitoring is completed. When the oven is heated to
the preset temperature, the system automatically shifts from continuous heating
mode to heat preservation mode. This setting is only applicable to DP260 and
DP260s.
Reagent heating temp: Set the range from 25~40
℃
, touch and pull the prompt
bar slide block to a proper value and then release. Click the “Setting” button, and
the setting of reagent heating temperature is completed. When reagent is heated
to the preset temperature, the system automatically shifts from continuous heating
mode to heat preservation mode. This setting is only applicable to DP260.
Exhaust fan mode: Includes three modes: normally open, intelligent and
normally closed. Users are advised to select the intelligent mode, and the system
decides whether to start or close the exhaust fan based on program running,
opening of transparent gate and opening of drawer.
Fans speed: includes three modes: high speed, moderate speed and low
speed. Users can set the rotation speed of exhaust fan according to use
environment or concentration value of waste gas.