V1.0
20170825
12
3. Apparatus
3.2 Apparatus Models
The apparatus is divided into three types according to the functional modules
and configuration of stations in the staining cabins: DP260, DP261 and DP260s.
Featuring the most complete configuration, DP260 has 1 loading station, 4 water
washing stations, 17 reagent stations (including 4 reagent stations with heating
functions), 1 oven station, and 3 unloading stations. Featuring simple configuration,
DP261 has 1 loading station, 4 water washing stations, 18 reagent stations
(without reagent heating function; the oven station is changed into reagent station),
and 3 loading stations. DP260s has 1 heating station, 4 water washing stations,
17 reagent stations (without reagent heating function), 1 oven station and 3
unloading station.
3.3 Technical Parameters
Conventional
Rated voltage: 220V
Rated frequency: 50Hz
Maximum power consumption: 800VA
GB 4793.1 classification: Electric shock protection Class I
Overvoltage category Class II
Pollution grade Class II
Non-continuously running equipment
Operating temperature range: 15 ℃ to 40 ℃
Relative humidity: 10% -80%, non-condensing
Size and weight
Host (L*W*Hmm): 1066 x 665 x 580
Net weight (without package weight): About 129 kg
Packed weight: About 148 kg
Capacity parameters
Sample glass slide throughput:
Capacity of slide stand:
Loading capacity:
Total number of stations:
Total number of reagent stations:
Reagent tank capacity:
Fixed memory capacity:
Dependent on the procedure structures
30 pieces of sample glass slide
At most 11 slide stands (running11
different staining programs)
26 (at most one of which is oven station
and 4 are water washing stations)
24 maximally (setting of station change)
480ml ± 20ml
≥100 two programs, each with up to 40
program steps.