5. Use
5.8 Staining
Before staining program is started, the following steps must be prepared:
Optimize all menu (configuration interface) parameters.
Reagents must be configured for stations, and are injected according to
this configuration.
Ensure that the program required by the staining can be started at any
time.
5.8.1 Loading slide stands
•
Slide stands can be loaded through the loading
station.
•
Open the drawer, and insert slide stands at the
loading station.
•
Close the drawer.
Pay attention to the operation sequence at the initial phase of oven
station.
The filing of reagent is forbidden during heating, or the heating operation
is forbidden after reagent filling at the oven station. The highest
temperature of oven can be 70
℃
. Reagent may catch fire, damaging the
apparatus and samples.
In order to correctly recognize slide stands, the
drawer must be opened for loading, and then
closed again. Do not open the transparent gate to
insert slide stands.
Fig.55