5. Use
5.6.2 Management of reagent changing
•
Click
“
Change Reagent
”
.
•
Click the inverted triangle arrow
on the right side of the
“
Change
Reagent
”
, and the reagent list
option box is displayed. Touch
to select the name of the
replacing reagent.
•
If the name of a reagent is not
available in the reagent list box,
the operator can click the
“
+
”
utton to manually add it.
•
Click
“
Save
”
and the reagent
changing can be completed.
It is not necessary for an operator to manually input a station name, time of change
and change. The system can recognize them according to the logging-in user
and current time.
When the reagent consumption of a station exceeds the preset maximum number
of slide stands or maximum use days, the top of the station on this interface will
be displayed in red, prompting the user to change the reagent. For the details,
please refer to Section 5.5.4.
Please make sure that the reagent in the reagent tank is consistent with the name
of the reagent configured on the reagent management interface; otherwise,
damage of specimen or poor staining may be caused.
The operation of changing reagent must be executed before the creation of
staining program. After reagent changing, if it is necessary to run the staining
program created before the changing, the program must be modified so that it
matches the reagent and station.
After program running, the operation of reagent changing cannot be executed.
The system pops out a warning dialog box, prompting the user to try later.