5. Use
Maintenance
Manual draining: Click the button “Manual draining”, and the started water pump
can discharge liquid in the reagent temperature-preserving tank to avoid
bacterial growth and apparatus erosion.
Moto test: Click the button “X, Y and Z Axis Test”, and the system executes the
shifting procedure at the corresponding axis to inspect whether the axial
movement of mechanical arm is intact.
Rinse station test: Click the button “Testing of Water Inlet at No. 5, 6, 7 and 8
Stations”, and the system executes the washing procedures of corresponding
water washing stations to ensure that the water inlet functions at the W5, W6, W7
and W8 stations are good.
Drainage test: Click the button “Drainage test”, and the W5, W6, W7 and W8
stations start water inlet simultaneously to check for blocking. The water inflow
lasts 5 minutes.
Oven test: Click the button “Oven Test”, and the system starts the oven for
heating to inspect whether the temperature rise is normal.
Fan test: Click the button “Fan test”, and the system starts the exhaust fan to
check whether the fan works normally.
Robotic arm step by step: Click the button “Execute” after the names of
corresponding stations are input in the columns of initial stations and target
stations. The, the mechanical arm catches slide stands at the two stations to
examine the accuracy of locating.