5. Use
Power-on Self-test: a program of self-inspection on all hardware when the
apparatus is started. If this item is checked, the apparatus executes this program
during start-up. Self-inspection lasts about three minutes, and other operations
cannot be performed during self-inspection.
QC mode: started by default in the system. Users can touch the screen to
uncheck this item. If it is unchecked, the quality control setting of quality control
modules (refer to Section 5.5.4) is ineffective, and the system does not give
warnings by pop-up box.
Tank liquid level sensor: started by default system . Users can touch the
screen to uncheck this item. If it is unchecked, the system does not detect the liquid
levels of waste liquid tan and water washing tank.
Shield detection: started by default system. Users can touch the screen to
uncheck this item. If it is unchecked, the system does not detect whether to open
the protective cover.
Add the reagents that need to wait in the first step of the program: When the
device is running multiple times, the robot is busy or the site time conflicts The new
shelf can not be loaded and causes the dye to wait for a long time at the loading
site. At this point can be set to the first step of the procedure need to wait for the
reagent (the system default optional reagent for the dip for too long and no effect
on the dye effect of reagents, such as xylene, alcohol, etc.), so the device will
automatically load the site waiting time Assigned to the first step of the reagent in
the dip time, so as to avoid the dye to kill.
Automatic program: Closed by default in the system. Users can touch the
screen to check this item. If it is checked, the system automatically executes the
staining program at the P1 shortcut station when it detects a slide stand at the
loading station and the user cannot execute other operations within the specified
period of time.
DP261 does not have the oven and reagent heating function. After ex-factory
configuration, operations cannot be performed to set oven and water tank temperature.
DP260s does not have reagent heating functions. After ex-factory configuration,
operations cannot be performed to set water tank temperature.