![CareFusion IVAC P6000 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/carefusion/ivac-p6000/ivac-p6000_technical-and-service-manual_2557183009.webp)
Introduction and Start Up
P7000, P6000, TIVA, TCI & TIVA
9/106
1000SM00012 Issue 8
Guide to handling the pressure disc
1. Insert the pressure disc into pressure disc support slots as shown in Figure 1.
2.
Push pressure disc securely home as shown in Figure 2.
When removing the pressure disc pull it with your finger inside the disc recess as shown in Figure 3. above.
Loading a Syringe
(continued)
1. Connect pump to AC Mains.
Starting the Pump
Figure 1.
Figure 2.
Figure 3.
Pressure disc support slots
Pressure disc
Disc recess
2. Press the
A
button to switch pump ON (Mark II), or switch on power switch on side of pump (Mark I).
3.
CLEAR DATA?
NO
retains previous data.
YES
clears previous infusion data.
4. Load syringe. See instructions in previous section.
5. Insert pressure disc into pressure transducer (Model P7000 only). See 'Guide to handling the pressure disc' above.
6. Confirm syringe.
7. Change the rate if necessary using the
NO
keys.
8. Purge: Press the
G
button followed by the
PURGE
softkey.
9. Connect the pump to test equipment as required (see Chapter 2, 'Configuration & Calibration' and Chapter 3, 'Routine
Maintenance').
10. Press the
H
button to start operation.
Important: To verify continued operation on battery power, prior to beginning an infusion, remove
the pump from the AC mains, confirm the pump continues to function on battery power and then
reconnect the pump to the AC mains.
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...