![CareFusion IVAC P6000 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/carefusion/ivac-p6000/ivac-p6000_technical-and-service-manual_2557183044.webp)
Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 44/106
1000SM00012 Issue 8
Accessing the Pump
Procedure
1. Remove the six case retaining screws and washers located on the base of the pump.
2. Carefully separate the upper and lower case halves and disconnect cables.
3. Reassemble in reverse order. Note that when closing the case halves, ensure the grey ribbon cable is stowed in front
of the PSU heatsink.
Ensure the pump is disconnected from the AC power supply and switched off before attempting to service the
pump
The pump contains static-sensitive components. Observe strict ESD precautions at all times
Always protect the plunger holder and syringe clamp when the pump is upside down. For regular servicing,
the use of the case support cradle, part number 0000JG00004, is recommended
Ensure that no undue force is applied to the plunger holder and the leadscrew, when the unit is placed upside
down to remove the six case retaining screws on the base
Batteries should be disposed of as outlined by the local country regulations: do not send back to the
manufacturer
For fastener torque settings, refer to Appendix C, 'Fitting & Replacement Guidelines'
Only use CareFusion recommended spare parts
Following all spare part replacement and repair activities, testing must be performed in accordance with the
Performance Verification Procedure (PVP), see Chapter 3, 'Routine Maintenance'.
(
D)
Foot (x4)
(
F
) Upper Case
(
E
) Lower Case
(
A/B
) Case screws (x6) / Washers (x2)
(
C
) Washers (x4)
Spare Parts
Item
Description
Part Number
A
SCREW M4 X 50 PAN POSI
0000ME00302
B
WASHER M4 WAVEY SST
0000ME00045
C
WASHER M4 PLAIN ZINC PLATED
0000ME00310
D
SPARE UPGRADE MOULDED FOOT
1000SP01066
E
KIT CASE LOWER TIVA/TCI
6000SP00047
E
SPARE CASE LOWER P7000 MKI
7000SP00010
E
SPARE CASE LOWER P7000 MKII
7000SP00027
F
UPPER CASE 6000
6000ME00010
F
SPARE CASE UPPER TCI
6000SP00045
F
SPARE KIT UPPER P6000/TIVA
6000SP00046
F
SPARE UPPER P7000 MKII
7000SP00033
*
BLANK ON/OFF SWITCH LOWER (lower case side)
6000ME00006
* item not shown
The information in this chapter is based on the Mark II pump unless otherwise stated.
Introduction
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...