Configuration and Calibration
P7000, P6000, TIVA, TCI & TIVA 18/106
1000SM00012 Issue 8
General Options (251)
(continued)
GENERAL OPTIONS MENU (Mode
l
P6000/Mode
l
P7000)
Software version: V3R2
Option
Description
P6000* P7000*
AUTOSAVE
Enable/disable Autosave feature (option to retain previous settings when pump is
switched on). Enabled: Confirmation is requested at power up to clear or to use previous
patient data. Disabled: Patient information is cleared on power up.
BACKOFF
Enable/disable Backoff feature (feature where pump action reverses to release infusion
system pressure build-up and minimise post occlusion bolus following a pressure alarm).
Enabled: Backoff feature ON. Disabled: Backoff feature OFF.
FAST START AND
BACKOFF
Enable/disable Fast Start and Backoff features (Faststart feature - drive slack fast start
at start of an infusion, Backoff feature - see above). Enabled: Fast Start and Backoff ON.
Disabled: Fast Start and Backoff OFF.
SET VTBI OVER TIME
Enable/disable Volume to be Infused (VTBI) over time option (allows the setting of a
fixed volume to be delivered over a fixed period of time). Enabled: VTBI OVER TIME
option available via the
D
button. Disabled: VTBI OVER TIME option not available.
VTBI
Enable/disable the Volume to be Infused (VTBI) option (allows the setting of a fixed
volume to be delivered). Enabled: VTBI option available via the
D
button. Disabled: VTBI
option not available.
INDUCTION
Enable/disable the INDUCTION option (option to set an induction volume to be
delivered). Enabled: INDUCTION option available via the
D
button. Disabled: INDUCTION
option not available.
MULTIDOSE
Enable/disable the MULTIDOSE option (feature allowing the delivery of a prescribed dose
to be repeated over a specific period). Enabled: MULTIDOSE option available via the
D
button. Disabled: MULTIDOSE option not available.
RATE LOCK
Enable/disable RATE LOCK option (anti-tamper feature preventing rate changes, bolus/
purge operations and pump powerdown). Enabled: RATE LOCK option available via the
D
button. Disabled: RATE LOCK option not available.
QUIET MODE
Enable/disable Quiet Mode feature (keypress tones, low priority alarms and power down
sequence tones are muted). Enabled: Quiet mode feature ON. Disabled: Quiet Mode is
OFF.
PRESSURE ALARM
Set pressure alarm value between 1mmHg and 750mmHg. Default is 300mmHg.
MAX. PRESSURE ALARM
Set maximum pressure limit value between 1mmHg and 750mmHg. Default is
750mmHg.
AUTO PRESSURE ALARM
Enable/disable Auto Set Pressure (AUTO) option (option to adjust the pressure alarm
level by a set pressure value above the measured in-line pressure, 15 mins. after start of
infusion). Enabled: AUTO option available via the
F
button. Disabled: AUTO option not
available.
AC FAIL ALARM
Enable/disable AC Fail alarm (activated when AC power disconnected and pump
operating on battery power). Enabled: AC Fail alarm ON. Disabled: AC Fail alarm OFF.
RATE TITRATION
Enable/disable ability to adjust (titrate) infusion while infusion is running. Enabled: rate
can be changed while the pump is infusing, without putting pump on hold. Disabled:
START button must be pressed to confirm new rate when rate changes are made while
infusing using the rate adjust keys.
NEOI WARNING
Set the Near End Of Infusion (NEOI) warning time between 1 min. and 15 mins. (NEOI
warning signals the syringe is almost empty allowing time to change syringe).
EOI POINT
Set the End Of Infusion (EOI) volume as a percentage of the syringe volume between
0.5% and 2.0%.
KVI at EOI
Enable/disable Keep Vein Open at End of Infusion (KVO at EOI). Enabled: Pump will
switch to run at the KVO rate 1.0ml/h (or the current set rate if this is lower) at the EOI
point. Disabled: Pump will stop at the EOI point.
BEAM ALARM
Set the default beam occlusion alarm level between level 0 (L0) and level 7 (L7).
WEIGHT
Set a default patient weight between 0.1kg and 150kg. Factory default is 70kg.
PURGE RATE
Set the purge rate volume between 100ml/h and 500ml/h. (Rate used during PURGE
operation).
BOLUS
Enable/disable Bolus function. (Function which administers a controlled volume of fluid
or drug at an increased rate). Enabled: Bolus function allowed. Disabled: Bolus function
not allowed.
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...