Configuration and Calibration
P7000, P6000, TIVA, TCI & TIVA 25/106
1000SM00012 Issue 8
If the internal RAM or its associated battery is replaced on the Control PCB, or if the pump fails with an 'ER5 RAM' error it
will be necessary to do the following:
Clear the internal RAM:
1. Power up the pump holding down the
H
button.
2. Enter the access code
611
.
3. Press
ENTER
and wait for the RAM to be cleared.
4. When the message
RAM CLEARED
appears, switch the pump off.
Fully calibrate pump. Perform each of the four calibration procedures, as described in this chapter.
Reconfigure the pump:
Set Configuration, drugs and protocol (it may be possible to use the Teach/Learn facility, see Chapter 3, 'Routine
Maintenance' for instructions).
Enter access code
251
and set time.
Enter access code
359
and set language.
Enter access code
376
and set service date.
Enter access code
901
to set to dedicated (Model P7000 only).
Carry out Performance Verification Procedure (PVP). See Chapter 3, 'Routine Maintenance' for instructions.
Clearing Internal RAM (611)
Warning: Do not clear the RAM unless absolutely necessary, because all calibration and
configuration in the pump will be cleared.
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...