Routine Maintenance
P7000, P6000, TIVA, TCI & TIVA 27/106
1000SM00012 Issue 8
The self-test procedure is designed to allow confirmation of many of the pump functions, defaults and calibrations
without requiring internal inspection.
1. Enter the access code
123
. See 'Entering an Access Code' in Chapter 2 for instructions.
2. The pump now proceeds through a series of tests. Press the
NEXT
button to move to the next test.
Refer to the table below for details of each test.
Important
: If the pump fails the test sequence at any stage, it should be taken out of service and inspected by a qualified
service engineer.
For routine maintenance, the following tests and performance verification procedures should be performed in addition
to the tasks described in the section 'Physical Inspection and Clean'.
Refer to the relevant
DFU
for the recommended routine maintenance period.
Self-Test Procedure (123)
Test
Display
Description/Action
Review software setup
SOFTWARE REVISION
PROGRAM CRC
SERIAL NUMBER
LANGUAGE
TCI LANGUAGE (TCI & TIVA
only)
Displays software version, program CRC, serial number and language.
Review calibration data
setup
SYRINGE REVIEW
Displays syringe, beam and linear calibration figures. Pressure
calibration shown on P7000 only.
Internal PSU and mains
voltage
PSU VOLTS: _ . _ _ V
REG VOLTS: _ . _ _ V
PSU VOLTS: On battery supply typically: 5.50V - 6.50V
PSU VOLTS: On mains supply: 6.50V - 7.50V.S
Audible Alarms
AUDIO: ALARM
Check loud alarm sound beeps.
AUDIO: WATCHDOG
Check loud alarm sounds continuously.
AUDIO: KEYPAD
Check soft alarm sounds continuously.
Visual: Display
DISPLAY TEST
Check that the display displays an even graduated grey tone.
Visual: Backlight
BACKLIGHT FULL / DIM / OFF
The display starts the test on FULL backlight, goes DIM and then turns
OFF. At the end of the test it turns back on to FULL.
Visual: LED Indicators
LEDS: FLASHING
Check that the STOP, START and BATTERY LED’S are flashing. Note: the
AC LED does not flash, it remains constant.
Touch panel buttons
KEY TEST: 00
Press buttons in turn from START (01) to PURGE (13).
H
01
C
09
I
02
E
or
F
10
N
03, 04
A
or
J
11
O
05, 06
D
12
C
07
G
13
C
08
Declutch switch
DECLUTCH: 1
Squeeze plunger holder finger grips and check that the display
alternates between 1 (engaged) and 0 (disengaged - finger grips
squeezed together).
Plunger detector
SYRINGE PLUNGER: 0
Press plunger plate button and check display switches from 0 (No
syringe fitted) to 1 (Syringe plunger fitted).
Motor encoder
MOTOR / ENCODER : 1
Motor is pulsed while encoders are tested. Motor moves forwards and
backwards as encoders pass.
Linear potentiometer
LINEAR POT : _ _ . _ _V
Declutch the transmission and slide plunger holder to the far left, check
the value displayed (approx. 00.19V). Declutch and slide the plunger
holder to the far right and check the value displayed (approx. 03.00V).
Syringe size detection
SYRINGE POT : _ _ . _ _ V
Lift the syringe clamp and check that the values displayed increase
within the normal range (approx. 0.05V and 3.00V).
Introduction
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...