Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 53/106
1000SM00012 Issue 8
Upper Case Assembly
Replacing the upper case
1. To replace an upper case, it will be necessary to fully strip down the old case and insert all the components into the
new upper case. This task requires a good knowledge of the pump, so be certain that you are fully conversant with all
the procedures in this chapter before undertaking this replacement.
2. To strip down each upper case subassembly, follow the instructions in the relevant section of this chapter. These
sections are:
Accessing the Pump
Control PCB, Display PCB
Transmission Assembly Removal
Transmission Assembly Breakdown
Syringe Size Pot, Syringe Clamp
Aerial Assembly (Model TCI & TIVA)
Pressure Disc Holder (Model P7000)
Pressure Transducer (Model P7000)
Display, Keypad
Labels
3. To reassemble the components into the new upper case, simply reverse the order of disassembly.