![CareFusion IVAC P6000 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/carefusion/ivac-p6000/ivac-p6000_technical-and-service-manual_2557183046.webp)
Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 46/106
1000SM00012 Issue 8
Battery
Procedure
1. Disconnect the battery cable from the Power Supply Unit (PSU).
2. Remove the two screws and washers which secure the battery restraining plate. Disconnect earth cable (Model TCI &
TIVA only).
3. Remove the restraining plate and foam pad.
4. Lift out the battery.
5. Reassemble in reverse order.
Spare Parts
Item
Description
Part Number
A
BATTERY 6V SLA RECHARGE
0000EL00004
B
FOAM PAD BATTERY
1000ME01064
C
PLATE BATTERY RESTRAINT PUNCHED
1000ME01123
D
SCREW M3X6 PAN HD POSI ZP+P
0000ME00221
E
WASHER M3 WAVEY SST
0000ME00015
F
LABEL SET P6000/TIVA/P7000 MKII
1000LB01449
F
LABEL SET TIVA TCI
6000LB00015
G
ASSY CABLE BATTERY
1000SP00009
H
EARTH CABLE PE TO BEAM
6000SP00006
I
EARTH CABLE PE TO BATTERY PLATE
6000SP00005
(
H
) Earth cable
(
I
) Earth
cable
Mode
l
TCI & TIVA
(
G
) Battery cable
(
A
) Battery
(
C
) Restraint plate
(
B
) Foam pad
(
D/E
) Screws (x2) /Washers (x2)
(
F
) Label (part of label set)
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...