Specifications
P7000, P6000, TIVA, TCI & TIVA 79/106
1000SM00012 Issue 8
The IVAC® P7000 Syringe Pump is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled.
The user of the IVAC® P7000 Syringe Pump can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the IVAC® P7000 Syringe Pump as recommended
below, according to the maximum output power of the communications equipment.
Rated Maximum Output
Power of Transmitter
W
Separation Distance According to Frequency of Transmitter
m
150 kHz to 80 MHz
Outside ISM bands
3.5
d
= [------]
√P
V
1
150 kHz to 80 MHz
In ISM bands
12
d
= [------]
√P
V
2
80 MHz to 800 MHz
12
d
= [ ------]
√P
E
1
800 MHz to 2.5 GHz
23
d
= [------]
√P
E
1
0.01
0.03
0.12
0.12
0.23
0.1
0.11
0.38
0.38
0.73
1
0.35
1.20
1.20
2.30
10
1.11
3.80
3.80
7.28
100
3.50
12.00
12.00
23.00
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be determined using the
equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1—At 80 MHz and 800 MHz, the separation distance for the higher frequency range apply.
Note 2—The ISM (Industrial, Scientific, and Medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957
MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Note 3—An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM frequency bands between
150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could
cause interference if it is inadvertently brought into patient areas.
Note 4—These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people.
Electromagnetic Compatibility (Model P7000)
continued
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...