Routine Maintenance
P7000, P6000, TIVA, TCI & TIVA 29/106
1000SM00012 Issue 8
Teach Learn
A pump can be configured/reconfigured by transferring data from one pump (Teacher pump) to another pump (Learner
pump) via the serial port.
Equipment required
Two pumps, of the same model (for example, two Model
P7000 pumps). Note that both pumps must have the same
version of software
RS232 cable. For Models
P6000/P7000/TIVA use part number 1000SP01008. For Model
TCI & TIVA use part number
6000SP00012
Procedure
1. Connect the Teacher pump to the Learner pump using the RS232 cable.
2. Switch both pumps ON.
3 On the Teacher pump, enter the access code
168
and on the Learner pump, enter the access code
167
.
The Teacher pump displays
TEACHING
and the Learner pump displays
LEARNING
.
4. When complete, select
OK
.
Possible reasons for Teach/Learn failure:
Pumps are different models
Software versions are different
Loose cable connection or faulty RS232 cable
Using Teach Learn to reconfigure a pump only transfers certain configuration settings. The
following settings are
not
transferred and must be reset manually: CLOCK SET (251), SET
LANGUAGE (359), SERVICE DATE (376), DEDICATED MODE (901/902), ENABLE UNITS (251) on
P7000 only.
Once Teach Learn is complete, all configuration settings on the Learner pump, including DRUG
SET UP, must be checked against the Teacher pump (original source).
Event Log Download
A PC application known as the Event Log Download Utility (ELDU) (part number 1000SP00209) is available to download
logs from IVAC
®
Syringe Pumps.
ELDU Operation
1. Click on
ELDU
icon on PC.
2. Click
Accept
to agree with Restrictions of Use and continue.
3. Select
Configure
from drop-down menu.
4. Select
Setup Pump
and choose
IVAC®
as pump type.
5. Select
Settings
to select log to be downloaded.
6. Select
Communications
then check options are set as follows:
- Required PC com port selected.
- Character mode and parity match IVAC
®
pump configuration.
- Delay mode is Normal.
- Pump address matches
COMMS PUMP ADDRESS
in
GENERAL OPTIONS
.
7. Click
OK
to confirm.
8. Connect the RS232 cable.
9. Power up pump.
10. Click
Download Log
from main PC screen.
11. Press
Close
, when finished.
12. Select
File
from drop-down menu and save file. Log may be printed here as required.
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...