Configuration and Calibration
P7000, P6000, TIVA, TCI & TIVA 19/106
1000SM00012 Issue 8
General Options (251)
(continued)
GENERAL OPTIONS MENU (Mode
l
P6000/Mode
l
P7000)
Software version: V3R2
Option
Description
P6000* P7000*
DEFAULT BOLUS
Set the default bolus rate value between 10ml/h and 1200 ml/h.
CAP BOLUS RATE
Set a maximum permissible bolus rate value between 10 ml/h and 1200ml/h.
CAP RATE
Sets maximum permissible infusion rate value between 1ml/h and 1200ml/h.
DISPLAY EVENT LOG
Enable/disable EVENT LOG option on OPTIONS menu. (Event Log shows record of timed
and dated pump events such as power on/off, alarms, rate changes).
Enabled: EVENT LOG option available via the OPTIONS menu (
D
button). EVENT LOG
option not available via the OPTIONS menu (
D
button). Note that the EVENT LOG
option is always available via the SERVICE ACCESS menu.
LOG LAST PATIENT ONLY
Enable/disable display of all patient data in Event Log. Enabled: Event Log displays last
patient data only (since new patient selected). Disabled: Event Log shows all patient
data.
NURSE CALL
Enable/disable the Nurse Call feature (hardware feature which allows the pump to
communicate with the hospital's nurse call system, typically linked to central nurse's
station). Enabled: Nurse Call Hardware enabled. Disabled: Nurse Call hardware disabled.
NURSE CALL INVERTED
Enable/disable inversion of Nursecall hardware output. Enabled: Nurse Call hardware
output is inverted. Disabled: Nurse Call hardware output normal.
COMMS PUMP ADDRESS
Set pump address for use in communications. Range is 1-254.
COMMS MONITOR ONLY
Enable/disable pump remote control via the comms link. Enabled: Control of pump
via monitor only. Disabled: Remote control of pump via comms link allowed. Note that
monitor control is always possible.
COMMS ODD PARITY
Enable/disable odd communication parity bit generation.
COMMS ASCII
Enable/disable ASCII communications mode.
Key:
= available option
= unavailable option
*
For pumps with software versions earlier than version V3R2, the options may vary, or will not be available.
Refer to the relevant IVAC® Syringe Pump
DFU
for comprehensive information.
Note: For default settings, refer to Appendix D, 'Configured Options and Drug Protocol Records'.
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...