Introduction and Start Up
P7000, P6000, TIVA, TCI & TIVA 10/106
1000SM00012 Issue 8
Pressure Level
Press the
E
button (On Model
P7000, press the
F
button). Line pressure occlusion alarm level
and current pressure level are shown on a graph. Use the
M
keys to adjust alarm level.
Purge/Bolus
Press the
G
button. If required, use the
M
buttons to set bolus dose required then press
and hold down the two
PURGE/BOLUS
softkeys together to deliver.
During PURGE/BOLUS, the pressure limit alarms are temporarily increased to maximum levels.
24 Hour Log
Press the
J
button. The 24 hour log of volume infused is shown. 24 Hour Log is also available
via the
D
button.
Option Button
Press the
D
button to display optional features:
DRUGS AND DOSING (or DRUG
NAME)
Allows the pump to be set up for use with a specific
drug and/or dosing protocol. Follow Drugs and Dosing
instructions as per relevant DFU.
SET VTBI*
Set a specific volume to be infused and set rate at end of
VTBI (not available on Model
TCI & TIVA or Model
TIVA).
SET VTBI OVER TIME*
Specify a VTBI and delivery time (not available on Model
TCI & TIVA or Model
TIVA).
SET BY DOSERATE*
Set rate in doserate increments (mg/kg/h). The flow rate
will be calculated from the doserate.
SET BY ml/h*
Set rate in flow rate increments (ml/h). The doserate will be
calculated from the flow rate.
CLEAR VOLUME
Clear the displayed volume infused.
INDUCTION*
Set an induction volume (P7000 only).
MULTIDOSE*
Set multidose volumes (P7000 only).
RATE LOCK*
Prevent rate being changed once infusion has started.
24H LOG
Shows volume infused over 24 hours.
EVENT LOG*
Displays the event log. Holds up to 800 individual events.
* these options are not displayed/available when feature is disabled.
Note that option names may vary and the Model
TCI & TIVA and Model
TIVA pumps have
additional model-specific features. For additional information, refer to the relevant
DFU
.
Basic Features
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...