Routine Maintenance
P7000, P6000, TIVA, TCI & TIVA 34/106
1000SM00012 Issue 8
Performance Verification Procedure
Model / Serial Number:
Service Order / Inventory Number:
Hospital Name / Reference:
Software Version:
Inspection
Physical inspection and clean
UPDATES
Mandatory when serviced
UPDATE
REF:
Fitted
Not fitted / Not
Applicable
Update P7000 with V3R2 software
TSM ch3
Update TIVA and TCI & TIVA with V3R2 software
TSM ch3
Inspect and fit display spacers on pumps manufactured prior to July 1999
with serial numbers below: 6001-00001 to 6001-13816, 6002-00001 to 6002-
11388, 6003-00001 to 6003-01126, 7001-00001 to 7001-13717
TSM ch6
Recommended when serviced
UPDATE
REF:
Fitted
Not fitted / Not
Applicable
Update P6000 with V3R2 software
TSM ch3
Bond the syringe clamp assembly on pumps manufactured prior to July 1999
with serial numbers: 6001-00100 to 6001-00972, 6002-00100 to 6002-00259,
7001-00106 to 7001-04432
TSM ch6
SERVICE
LOG
CH3
Check/set serial number and set service date (optional) (
376
)
Self Test
CH3
Check all functions in self-test (
123
)
Check and adjust time and date as required (
251
)
INFUSING
Alarms functionality check
Drive Disengaged, VTBI done, Syringe Location, Plunger Location, AC power fail, Set Removed (P7000), Near End of
Syringe, Syringe Empty, KVO
Ensure pump works on battery and AC mains
Verification
Tests
CH3
Linear speed test
Pump set to 100 ml/h, syringe type BD Plastipak 50, for a distance of 15 mm. 4
mins 52 secs to 4 mins 58 secs
______ mins ______ secs
Occlusion test
Pump set to 100 ml/h, syringe type BD Plastipak 50
Semi Dedicated, Occlusion alarm level L-3, 2.2kgf to 3kgf
OR Dedicated (P7000), alarm level 220mmHg, 3.1kgf to 3.9kgf
_____________ kgf
Line pressure test (P7000)
Alarm set to 50mmHg, pump alarms 27mmHg to 73mmHg
Alarm set to 500mmHg, pump alarms 455mmHg to 545mmHg
_____________ mmHg
_____________ mmHg
TCI function test Fit TCI syringe (Diprivan 1%). Set TCI mode: age=40, wt=70 kg,
TCI target=4.0 µg/ml. Elapsed time to reach 4.0 µg/ml=24 ± 1 sec.
______ mins ______ secs
Setup
Set rate to zero (or lowest value possible), Clear Volume Infused and VTBI
Clear Error / Alarm / Battery logs (as required)
Electrical
Safety Tests
Class II Type CF - P7000, P6000, TIVA
Class I Type CF - TCI & TIVA
Test in accordance with the standard
EN 60601-1 and test equipment
operation manual.
Test results are stored:
Electronically
¨
Print-out
¨
Other
¨
___________________
PASS / FAIL
Verification
Performed
By
________________________
Sign
____________________
______________________
Date
CHx indicates the chapter number in the Technical Service Manual (TSM) - 1000SM00012.
E.G. CH3 = Refer to TSM Chapter 3. DFU = Refer to the relevant DFU.
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...