Fitting and Replacement Guidelines
P7000, P6000, TIVA, TCI & TIVA 89/106
1000SM00012 Issue 8
Torque Guide
(continued)
Pressure Disc Holder Assembly (Model P7000 only)
Stage Description
Component Description
Qty
Established Process
Torque
Disc holder to Upper Case
Screw M3x8 Csk Hd Posi SS
4
20cNm
Final Pump Assembly
Stage Description
Component Description
Qty
Established Process
Torque
Lower Case to Upper Case
Screw M4x50 Pan Posi
4
Hand tight
Pole Clamp Fitting - Setting Adjustable Clamp
Stage Description
Component Description
Qty
Established Process
Torque
Fully loosen locknut with clamp secured
to Lower Case
10mm Nylok nut
N/A
N/A
Close clamp
Fit and torque
Alternatively, secure clamp to pole
and apply torque until force to move
pump is at desired level
5mm Allen headed bolt
N/A
10Nm
Set clamp movement force
Hold 5mm Allen headed bolt in exact
position set during previous stage
Simultaneously torque locknut to
ensure bolt maintains current level of
torque and position
Locknut
N/A
20Nm
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...