Disposal
P7000, P6000, TIVA, TCI & TIVA 100/106
1000SM00012 Issue 8
Removal Procedure
1. Remove the six case retaining screws and washers located on the base of the pump.
2. Carefully separate the upper and lower case halves and disconnect cables.
3. Disconnect the cables from the Control PCB.
4. Remove the five PCB fixing screws and washers and withdraw the Control PCB and Display PCB together.
5. Pull out the four standoffs/pins to separate the Control PCB and the Display PCB.
6. Desolder battery from the Control PCB.
The transmission is not shown here for
clarity—it is not necessary to remove the
transmission assembly in order to remove
the Control PCB and the Display PCB.
Battery
Battery Removal continued
Control PCB
PCB Screws (x5)