Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 56/106
1000SM00012 Issue 8
Transmission Assembly Removal
Spare Parts
Item
Description
Part Number
A
SCREW M4x20 CSK HD POSI SS
0000ME00255
B
SCREW M4x8 PAN HD POSI
0000ME00246
C
WASHER M4 WAVEY SST
0000ME00045
D
SPARE TRANSMISSION P6000/TIVA/P7000
7000SP00022
D
SPARE TRANSMISSION TCI
6000SP00027
E
SCREW M4x40 PAN HD POSI 2 ZP+P
0000ME00225
Procedure
1. Remove the four chassis screws and washers and two beam screws which secure the transmission assembly.
2. Disconnect cables and remove the transmission assembly from the upper case.
3. Reassemble in reverse order.
When refitting the transmission assembly into the upper case, ensure that the flexi circuit is adjusted
so the flex does not catch or click when mechanism is declutched or moved manually.
(
C
) Screws (x2)
(
E
) Screws (x4)
(
D
) Transmission assembly
(
A
) Beam screws (x2)
(
B/C
) Chassis screws (x3) Washers (x4)
Flexi circuit
(
D
) Transmission Assembly
(
E
) Chassis screw (x1)
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...