Configuration and Calibration
P7000, P6000, TIVA, TCI & TIVA 17/106
1000SM00012 Issue 8
General Options (251)
GENERAL OPTIONS MENU (Mode
l
TIVA/Mode
l
TCI & TIVA)
Software version: V3R2
Option
Description
TIVA*
TCI
&TIVA*
BACKOFF
Enable/Disable Backoff feature (feature where pump action reverses to release infusion
system pressure build-up and minimise post occlusion bolus following a pressure
alarm).
Enabled: Backoff feature ON (This does not apply when the pump is in TCI mode).
Disabled: Backoff feature OFF.
AC FAIL ALARM
Enable/disable AC Fail alarm (activated when AC power disconnected and pump
operating on battery power). Enabled: AC Fail alarm ON. Disabled: AC Fail alarm OFF.
NEOI WARNING
Set the Near End Of Infusion (NEOI) warning time between 1 min. and 15 mins. (NEOI
warning signals the syringe is almost empty allowing time to change syringe).
EOI POINT
Enable/disable Keep Vein Open at End of Infusion (KVO at EOI). Enabled: Pump will
switch to run at the KVO rate 1.0ml/h (or the current set rate if this is lower) at the EOI
point. Disabled: Pump will stop at the EOI point.
BEAM ALARM
Set the default beam occlusion alarm level between level 0 (L0) and level 7 (L7).
WEIGHT
Set a default patient weight between 0.1kg and 150kg. Factory default is 70kg.
PURGE RATE
Set the purge rate volume between 100ml/h and 500ml/h. (Rate used during PURGE
operation).
HANDS FREE BOLUS
Enable/disable Bolus function in a 'hands-free' way—no need to hold down buttons
during bolus infusion. (Bolus is a function which administers a controlled volume
of fluid or drug at an increased rate). Enabled: Hands Free Bolus function allowed.
Disabled: Hands Free Bolus function not allowed.
DEFAULT BOLUS
Set the default bolus value between 0.1ml and 25ml.
NURSE CALL
Enable/disable the Nurse Call feature (hardware feature which allows the pump to
communicate with the hospital's nurse call system, typically linked to central nurse's
station). Enabled: Nurse Call Hardware enabled. Disabled: Nurse Call hardware
disabled.
NURSE CALL INVERTED
Enable/disable inversion of Nursecall hardware output. Enabled: Nurse Call hardware
output is inverted. Disabled: Nurse Call hardware output normal.
COMMS PUMP ADDRESS
Set the communications address of the pump.
COMMS MONITOR ONLY
Enable/disable pump remote control via the comms link. Enabled: Control of pump
via monitor only. Disabled: Remote control of pump via comms link allowed. Note that
monitor control is always possible.
COMMS ODD PARITY
Enable/disable odd communication parity bit generation.
COMMS ASCII
Enable/Disable ASCII communications mode.
TCI
Enable/disable TCI option. Enabled: TCI option is activated (TCI hardware must be
fitted). Disabled: TCI option not activated.
DISPLAY TCI DECREMENT
TIME
If enabled, the display shows the TCI Decrement Time Icon (TCI hardware must be
fitted).
Key:
= available option
= unavailable option
*
For pumps with software versions earlier than version V3R2, the options may vary, or will not be
available. Refer to the relevant IVAC® Syringe Pump
DFU
for comprehensive information.
Note: For default settings, refer to Appendix D, 'Configured Options and Drug Protocol Records'.
1. Enter the access code
251
to display the
CONFIGURATION MENU
.
2. Select
GENERAL OPTIONS
using the
M
keys and press the
OK
softkey. The
GENERAL OPTIONS MENU
is displayed:
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...