Configuration and Calibration
P7000, P6000, TIVA, TCI & TIVA 14/106
1000SM00012 Issue 8
Enter the access code
251
(see 'Entering an Access Code' on the previous page for instructions). The
CONFIGURATION
menu is displayed:
CONFIGURATION MENU
Option*
Description
DRUG SETUP
Set drug names and protocols. See 'Drug Setup' on the following pages for instructions.
GENERAL OPTIONS
See 'General Options' on the following pages for further details.
CLOCK SET
Set the current date and time.
To set the clock, use
M
and
NEXT
softkey to adjust then
OK
to store.
HOSPITAL NAME
Enables hospital/ward name to be displayed during the power-up sequence.
To set the hospital name, use
M
to toggle through characters,
NEXT
softkey to go to next
character then
OK
to store.
ENABLE SYRINGES
Configure the type and size of syringes permitted for use.
To enable syringes:
1. Use
M
to select,
MODIFY
to enable/disable the selected syringe type then
OK
to store.
2. Turn the pump OFF and return to service or select the next
CONFIGURATION
option.
ENABLE UNITS
Configure the dose units permitted for use (Model P7000 only).
To enable units:
1. Use
M
to select,
MODIFY
to enable/disable the selected dose then
OK
to store.
2. Turn the pump OFF and return to service or select the next
CONFIGURATION
option.
Configuration Options (251)
*
Note:
For software versions earlier than version V3R2, the options may vary, or will not be available. Refer to the
relevant IVAC® Syringe Pump
DFU
for comprehensive information.
Dedication Options (901/902)
The Mode
l
P7000 pump can be operated in one of two modes:
Fully dedicated mode
Operating the pump in this mode will remind a user that a pressure disc must be fitted at the start of any infusion. In this
mode occlusion pressures are always displayed in
mmHg
. To set, enter access code
901
.
Semi-dedicated mode
Operating the pump in this mode will require a user that a pressure disc must be fitted only when drugs and dosing
features are used. When a pressure disc is not in use, pressure levels L-0 to L-10 will be displayed. To set, enter access code
902
.
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...