Configuration and Calibration
P7000, P6000, TIVA, TCI & TIVA 22/106
1000SM00012 Issue 8
Beam Occlusion Calibration (717)
Excessive force will damage the plunger mechanism. Do not apply more than 10 kgf
±0.05
kgf
to the plunger mechanism at any time.
To convert Kilograms of Force (kgf) to Newtons (N) multiply by 9.806650. For example 10 kgf =
98.07N.
Step 1
0kgf
Step 3
Calibration Procedures
(continued)
Step 2
4kgf
Calibration tools required: 0000TG00020 (or 0000TG00200) and 0000JG00014
Calibration procedure:
Enter the access code
717
.
Fit calibration tools and position plunger as shown in Steps 1-3 below, zero the gauge dial. At each step press
CAL
softkey when required calibration force is reached.
Notes
:
1) If
CAL
does not appear in display, check for correct positioning of tool. If calibration cannot be performed, repairs
to the pump may be necessary.
2) The calibration values shown on the pump displays are for illustrative purposes only.
Allow 30 seconds for pressure to stabilise before pressing
CAL
softkey.
Power down to complete the calibration sequence.
For best results, before pressing CAL softkey, strive for mid-range force value.
Содержание IVAC P6000
Страница 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Страница 35: ...Chapter 4 Troubleshooting In this chapter Introduction 36 Error Codes 36 General Fault Diagnosis 38...
Страница 87: ...Appendix C Fitting and Replacement Guidelines...
Страница 98: ...Appendix E Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...
Страница 105: ...Page Intentionally Left Blank...