These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
9.7.4
Recommended separation distances between external power
conductors and the Vivo 3
Rated maximum current in
conductor (A)
Separation distance (m)
50-60 Hz: d= I/2πH= I/188
1
0.005
10
0.05
30
0.16
For conductors rated at a maximum current not listed above, the recommended separa-
tion distance d in meters (m) can be estimated using the equation d=I/2πH, where I is
the maximum current rating of the conductor in amperes (A) according to the transmit-
ter manufacturer; H is the Vivo 3 immunity compliance level to electromagnetic fields
in the 50-60 Hz frequency span (30 A/m).
9.8
Delivery Settings
Alarm
Default Setting
Low Pressure
10 cmH
2
O
High Pressure
25 cmH
2
O
High Pressure
Off
High Minute Volume
Off
Low Breath Rate
Off
High Breath Rate
Off
Apnea
Off
Low Pulse Rate
Off
High Pulse Rate
Off
Low SpO2
90%
High SpO2
Off
Disconnection (of mask)
On
Rebreathing (blocked mask leakage)
On
Low EPAP
Off
High EPAP
Off
Alarm Sound Level
5
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2