4.2.20
Heated Circuit
This setting defines whether a heated patient circuit shall be used as patient circuit.
By default, this setting can be modified in Clinical mode only, but the Vivo 3 may be
configured to allow modification in home mode as well.
WARNING!
Read the section 2.7
before activating the heated circuit, to make sure all conditions are met and
considered.
Property
Description
Value range
•
On
•
Off
Default value
Off
Available in modes
All modes
4.2.21
Heated Circuit Level
This setting defines the heating level, if
Heated Circuit
is set to
On
.
By default, this setting can be modified in Clinical mode only, but the Vivo 3 may be
configured to allow modification also in home mode.
Property
Description
Value range
1 to 5
Default value
1
Setting resolution
1
Available in modes
All modes
62 Treatment Functions and Settings
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2