6.7.1
Using the Attachable Humidifier
WARNING!
Read the section 2.7
before using the Vivo 3 with the attachable humidifier.
WARNING!
Risk of Burns
If you remove and open the humidifier just after using the ventilator, ensure that you
don
’
t touch the heater inside the humidifier since it can be very hot.
WARNING!
The humidifier is for single patient use only.
NOTE
The ventilator can also be used with active stand-alone and passive humidifiers
The humidifier is intended to humidify the patient air. It is intended for non-invasive
use only.
About the Attachable Humidifier
The attachable humidifier is intended for non-invasive use only.
The humidifier requires a connection to the Mains power supply to work.
The information in the below table is applicable to the recommended breathing system
configuration, which is Attachable humidifier and Heated circuit.
Property
Value
Humidifier
classification
Category 2 humidifier according to ISO 80601-2-74
Rated Flow
20-50 l/min
Operating Conditions
+15°C to +35°C. Humidity: RH from 15% to 95%.
Gas leakage
< 0.2 l/min at 50 cmH
2
O
Max humidification
output
> 13 mg/l, tested according to ISO 80601-2-74
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2