Immunity test
Compliance level
Electromagnetic environment
—
Guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
The relative humidity should be at least
5%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power sup-
ply lines
±2 kV for input/out-
put lines
Mains power quality should be that of a
typical commercial, hospital and residen-
tial environment.
Surge IEC
61000-4-5
± 0.5, ± 1 kV, ± 2 kV
line to line
Mains power quality should be that of a
typical commercial, hospital and residen-
tial environment.
Power frequency
(50/60 Hz) mag-
netic field IEC
61000-4-8
30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical loca-
tion in a typical commercial, hospital and
residential environment.
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines IEC
61000-4-11
5% UT, 0.5 cycle
(multiple phase
analysis);
5% UT, 1 cycle;
70% UT, 25/30
cycles (50/60 Hz);
0% UT, 5 sec (50/60
Hz);
The Vivo 3 runs on internal battery dur-
ing voltage dips, short interruptions and
voltage variations on power supply input
lines.
NOTE
U
T
is the mains voltage prior to application of the test level.
WARNING!
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the Vivo 3, including cables specified by the manufacturer. Otherwise, degra-
dation of the performance of this equipment could result.
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual