2
Press and hold the
On/Off
button
until the
Starting Treatment
progress
bar is filled.
⇒
The Vivo 3 now performs a func-
tion test indicated by a short beep
and then starts the treatment.
Check that the self test is per-
formed successfully, this is indicated
by:
•
A short beep indicating suc-
cessful alarm signalling
•
The alarm LED first lights yel-
low, then red
•
The audio pause LED lights
yellow
•
In about a second, both LEDs
are turned off
If the function test beep absents, take
the Vivo 3 out of use and contact
your supplier of the Vivo 3 e.
6.4
Stop the Treatment
1
Press and hold the
On/Off
button
until the
Stopping Treatment
progress
bar is filled.
⇒
A request for a complementary
action is now displayed for stop-
ping the treatment. If no comple-
mentary action is taken within 6
seconds, the Vivo 3 reverts to nor-
mal operation.