Environmental Condition
Specification
Transport and Storage
Temperature
•
+5°C to +45°C (Maximum 90 days)
• –
25°C to +70°C (Maximum 30 days)
If stored in temperatures outside normal operation
conditions, let the Vivo 3 acclimate before taking
it to use.
CAUTION!
The Vivo 3 must not be stored in a
warm place, such as direct sunlight or
close to a radiator.
When the ventilator is brought from
minimum/maximum allowed storage
temperature, ensure that it is warmed
up/cooled down for one hour before
starting it.
Humidity
RH from 15% to 95%, non-condensing.
Ambient Pressure Range
70 to 106 kPa
This corresponds to ~ 315 m below sea level to
~3000 m above sea level
Ventilator ingress protection
IP 22
Mechanical ingress protection: protected from
touch by fingers and objects greater than 12.5 mm
Liquid ingress protection: The device withstands
dripping water(equivalent to 3 mm rainfall
/minute) when not tilted more than 15 degrees
from vertical. The protection has been tested for
10 minutes (2.5 minutes in every tilt direction).
CAUTION!
There is a silicone lid to protect the
USB, SD card, and communication
ports. The IP22 classification is appli-
cable only when this lid is in place.
However, the accessory box can be
connected with retained IP22 classifi-
cation, but then only the lower part of
the silicone lid can be opened.
Ensure that the silicone lid on the
back of the ventilator is closed when
no accessories are connected.
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2
Содержание Vivo 3
Страница 1: ...DOC 007232 EN EU Y 2 2019 09 03 Vivo 3 Clinician s Manual...
Страница 2: ......
Страница 8: ......
Страница 136: ......
Страница 137: ...Appendix Appendices 137 Doc 007232 Y 2 Vivo 3 Clinician s Manual Appendices...
Страница 140: ......