Filter
Maintenance Intervals (minimum)*
Air inlet filter, grey (coarse)
•
Wash: every week.
•
Replace: every year or when assigning
the Vivo 3 to a new patient.
Air inlet filter, white (fine)
•
Replace: every 4th week or when
assigning the Vivo 3 to a new patient.
* If the Vivo 3 is used in an environment with high grades of pollen or pollutions,
shorter intervals might be required.
1
Turn off the Vivo 3 and place it on a clean dust free surface.
2
Pull out the filter holder and
remove the filters.
3
If required by the interval or if visibly dirty, wash the grey coarse filter:
1. Wash the filter using warm water and a mild soap.
2. Rinse thoroughly.
3. Dry the filter by first squeezing it in a towel and then letting it dry in the air.
Do not wring the filter.
4. Check that the filter is undamaged and completely dry before reinstalling it.
4
When reinstalling the air inlet filters
in the filter holder: first install the
grey coarse filter, then the white
fine filter.
5
Reinstall the filter holder on the Vivo 3.
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual