Property
Description
Setting resolution
Below 10 cmH
2
O: 0.5 cmH
2
O
Above 10 cmH
2
O: 1.0 cmH
2
O
Available in modes
•
PCV+A
•
PSV
•
S
•
S/T
•
T
4.2.3
Expiratory Positive Airway Pressure (EPAP)
The EPAP setting defines the airway pressure during the expiration phase.
Property
Description
Value range
2 cmH
2
O to 20 cmH
2
O.
For inspiratory pressure settings below 22 cmH
2
O, the max-
imum value for EPAP is limited to the inspiratory pressure
minus 2 cmH
2
O.
Default value
5 cmH
2
O
Setting resolution
Below 10 cmH
2
O: 0.5 cmH
2
O
Above 10 cmH
2
O: 1.0 cmH
2
O
Available in modes
•
PCV+A
•
PSV
•
S
•
S/T
•
T
4.2.4
Continuous Positive Airway Pressure (CPAP)
The CPAP setting defines the pressure that will be applied to the airways.
Property
Description
Value range
4 cmH
2
O to 20 cmH
2
O.
Default value
10 cmH
2
O
Setting resolution
Below 10 cmH
2
O: 0.5 cmH
2
O
Above 10 cmH
2
O: 1.0 cmH
2
O
Available in modes
CPAP
54 Treatment Functions and Settings
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2