Procedure
Prerequisites
•
The Vivo 3 shall be connected to the power supply.
•
The Vivo 3 shall be turned on and in
Standby
mode.
1
Press and hold the
On/Off
button until the
Starting Treatment
progress bar is filled.
2
When starting treatment, an automatic alarm signal test is performed. Check that
the test is performed successfully, this is indicated by:
•
A short beep indicating functional audio signaling.
•
The alarm LED first lights yellow, then red, indicating functional visual
signaling.
•
The audio pause LED lights yellow.
•
In about a second, both LEDs are turned off.
If the test fails, do not use the Vivo 3. Contact your supplier of the Vivo 3 for a
technical check.
3
Perform a pre-use test (select
Others
—
>
Pre-use Test
).
4
Disconnect the power cord for more than 5 seconds.
Check that the device switches to the internal battery and that the information mes-
sage
Switched to Internal Battery
is shown on the screen.
If not, contact your supplier of the Vivo 3.
5
Reconnect the power cord.
Check that the Vivo 3 switches to the mains supply.
If not, contact your supplier of the Vivo 3.
5.6
Adjusting the Settings
WARNING!
Risk of Asphyxia or Personal Injury
Using incorrect settings may cause personal injury or severe medical conditions, such
as hypercarbia, producing arterial acidemia.
The therapy settings shall be based on a physician
’
s description. Changes to settings
must be made by authorized clinical personnel only.
For detailed information about treatment modes and settings, see 4
The Vivo 3 shall be turned on and in
Standby
mode.
Follow the instructions below when setting up the Vivo 3.
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual