4.2.12
Target Volume
The Target Volume setting defines the tidal volume that the Vivo 3 will aim for while
ventilating the patient in a pressure mode. This is done by adapting the Inspiratory Pres-
sure between two adjustable pressure limits: Min IPAP and Max IPAP.
Property
Description
Value range
100 ml to 1500 ml.
The available range for Target Volume might also be limited
by the set values for
Breath Rate
.
Default value
100 ml
Setting resolution
Below 500 ml: 20 ml
Above 500 ml: 50 ml
Available in modes
•
PCV+A
•
PSV
•
S
•
S/T
•
T
4.2.13
Max Pressure
The Max pressure setting is only used when Target Volume is activated.
Max pressure defines the upper pressure limit up to where the Vivo 3 can increase the
pressure to reach the set Target Volume. If Target Volume is not reached at Max pres-
sure, the Vivo 3 will continue to ventilate at this Max pressure setting.
Property
Description
Value range
4 cmH
2
O to 40 cmH
2
O.
Default value
Same as the value for
IPAP
Setting resolution
Below 10 cmH
2
O: 0.5 cmH
2
O
Above 10 cmH
2
O: 1.0 cmH
2
O
Available in modes
•
PCV+A
•
PSV
•
S
•
S/T
•
T
The air pressure is measured by a sensor, and software regulates the blower that gener-
ates the pressure.
58 Treatment Functions and Settings
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2