Rise time
9
Insp. trigger
9
Exp. trigger
3
Min insp. time
Off
Max insp. time
Off
Breath rate
12 bpm
Insp. time
2.0 S
All alarm settings shall be set to Off, if possible.
4
Start the treatment.
5
Set the
High Pressure Alarm
level to Off.
6
Set the
Low Pressure Alarm
level to 20 cmH
2
O.
⇒
The Low Pressure Alarm shall be given.
8.4.2
Optional Tests
In this procedure, methods for additional alarm tests are described. These tests are
optional and not needed to ensure safe use of the Vivo 3.
8.4.2.1 High Pressure Alarm
1
Connect the Vivo 3 patient circuit to a test lung and a CPAP device.
2
Set High Pressure alarm to 10 cmH
2
O.
3
Set the CPAP device treatment pressure to 12 cmH
2
O.
4
Adjust the Vivo 3 settings as follows:
•
Ventilation Mode: T
•
Target Volume: Off
•
IPAP: 10 cmH
2
O
•
EPAP: 5 cmH
2
O.
•
Breath Rate: 12 bpm
5
Start the treatment on both the Vivo 3 and the CPAP device.
6
Wait until the High Pressure alarm is activated.
7
Stop the treatment. The test is completed.
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual