Property
Description
Setting resolution
Below 10 cmH
2
O: 0.5 cmH
2
O
Above 10 cmH
2
O: 1.0 cmH
2
O
Available in modes
•
PCV+A
•
PSV
•
S
•
S/T
•
T
•
CPAP
4.2.18
Humidifier
This setting defines whether a humidifier shall be used.
By default, this setting can be modified in Clinical mode only, but the Vivo 3 may be
configured to allow modification in home mode as well.
WARNING!
Read the section 2.7
before activating the humidifier, to make sure all conditions are met and considered.
Property
Description
Value range
•
On
•
Off
Default value
Off
Available in modes
All modes
4.2.19
Humidifier Level
This setting defines the level of humidification, if
Humidifier
is set to
On
.
By default, this setting can be modified in Clinical mode only, but the Vivo 3 may be
configured to allow modification in home mode as well.
Property
Description
Value range
1 to 5
Default value
1
Setting resolution
1
Available in modes
All modes
Treatment Functions and Settings 61
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual
Содержание Vivo 3
Страница 1: ...DOC 007232 EN EU Y 2 2019 09 03 Vivo 3 Clinician s Manual...
Страница 2: ......
Страница 8: ......
Страница 136: ......
Страница 137: ...Appendix Appendices 137 Doc 007232 Y 2 Vivo 3 Clinician s Manual Appendices...