6.8.3
Transfer Data using an USB Cable
WARNING!
The PC must be placed outside the patient area (i.e. more than 2 meters from the
patient).
1
Connect the USB cable to the
Vivo 3.
Make sure it is fitted correctly.
2
Connect the other end of the cable to the PC.
3
Start the software for viewing or retrieving data.
6.9
Basic Troubleshooting
Problem
Action
The ventilator doesn
’
t start.
Check the connection of the power cord
between the ventilator and the power
outlet.
The ventilator starts, but the patient cir-
cuit doesn
’
t deliver any air.
Ensure that both ends of the patient cir-
cuit are correctly connected.
Straighten the patient circuit or replace it.
The humidifier doesn
’
t work properly.
If the humidifier is incorrectly assembled,
disassemble it and then assemble it
correctly.
If the air is dry despite using the humidi-
fier, increase the level of humidification.
If the humidifier doesn
’
t deliver any heat,
check that the Vivo 3 is powered from
the AC power supply; the humidifier
doesn
’
t work on battery.